Glow wire testing answers a simple question with a complicated procedure: if a hot metal element touches this plastic part, does the part keep burning after the element is pulled away, and does it drop flaming material onto the tissue below? Clause 13 of IEC 60335-1, clause 13.3 of IEC 60598-1 and the materials clauses of IEC 60695-2-11, -2-12 and -2-13 all depend on this answer.
This apparatus is built around the parts that actually influence the result: the glow wire tip geometry, the applied force, the approach speed and the temperature of the wire itself. Everything else, including the enclosure and the touch screen, exists to make those four things repeatable shift after shift.
| Standards | IEC 60695-2-10, IEC 60695-2-11, IEC 60695-2-12, IEC 60695-2-13, IEC 60335-1 clause 13, IEC 60598-1 clause 13.3, GB/T 5169.10 |
|---|---|
| Glow wire tip | Nickel chromium Φ4 mm, machined loop, replaceable |
| Applied force | 1.0 N ±0.2 N for standard testing, 0.3 N available for special clauses |
| Penetration | Limited to 7 mm by mechanical stop, depth adjustable |
| Approach speed | Motorised 10–25 mm/s, constant, no operator influence |
| Temperature range | Room temperature to 1000 °C, set point in 5 °C steps |
| Temperature sensing | K type thermocouple Φ0.5 mm in the wire loop, closed loop control |
| Application time | 30 s ±1 s, timer for after flame observation 30 s |
| Test chamber | Darkened stainless chamber, 1000 mm internal height, observation window |
| Specimen positioning | XY carriage with 3 axis adjustment, sample clamp with weights |
| Controller | PLC with 7 inch touch screen, 40 stored recipes, USB export of run log |
| Safety | Door interlock, exhaust fan, over temperature cut off, emergency stop |
| Power supply | AC 220 V 50/60 Hz, 2.5 kW (110 V available) |
| Footprint | Approx. 1200 x 900 x 1600 mm, weight 180 kg |
A glow wire tip that has been re-bent, cleaned with abrasive paper or overheated loses its geometry and its temperature curve. We supply the tip with a dimensional record and a thermocouple seat machined to a depth that keeps the junction inside the loop. Tip replacement is a two-minute job with a locating pin, so labs replace rather than repair.
Force is applied by a calibrated weight stack acting on the carriage, so when the sample softens the wire keeps 1 N instead of driving deeper. This is where manual rigs quietly fail: the operator pushes and the penetration stop hides the error.
The controller holds the set point at the tip, not at the transformer. A K type thermocouple Φ0.5 mm is welded into the loop and the PID loop compensates when the sample starts to draw heat. Without that, the tip can sag 30 to 50 °C during the first seconds of contact.
Every run writes a log with set temperature, measured temperature, force, application time, penetration and operator ID. The log exports to USB as CSV, which is what a reviewer asks for when a client disputes a GWFI value.
| Comparison item | HongCe glow wire apparatus | Manual entry level rigs | European reference brands |
|---|---|---|---|
| Wire temperature control | Closed loop at the tip, ±10 °C band typical | Transformer dial, drifts 30–50 °C under load | Closed loop, comparable |
| Approach motion | Motorised constant speed 10–25 mm/s | Hand lever | Motorised |
| Force application | Calibrated weight stack, 1.0 N ±0.2 N | Spring, force unknown | Weight stack |
| Run record | CSV log with temperature curve, USB export | Paper notes | Software licence, extra cost |
| Tip replacement | Locating pin, 2 minutes, tip in stock | Whole assembly replaced | Lead time weeks |
| Budget level | Mid range, around half of European list | Lowest | Highest |
No. Both procedures use the same apparatus with different temperature stepping logic. The recipe page keeps the two methods separate so an operator cannot mix them.
We supply a verification thermocouple and a procedure sheet. Most labs also send the unit for annual third-party calibration, and the certificate can be arranged through our calibration partners at extra cost.
Plain tissue paper with a mass between 12 and 30 g/m², laid on a 10 mm pine board at 200 mm below the sample, as described in IEC 60695-2-10. We ship a starter pack with the machine.
Yes, but the tip is a consumable. Running at 960 °C every day shortens tip life to roughly 3 to 6 months. Tips are inexpensive and stocked.
The chamber has an exhaust port and fan. You need to connect it to your lab extraction duct, and we provide the flange dimension in the installation drawing.
20 to 25 days for the standard unit, 30 days if a third-party calibration certificate is ordered together.
Guangzhou HongCe Equipment Co., Ltd, founded in 2003 and formerly known as HongTong Electromechanical, builds fire hazard, IP code and electrical safety test equipment in Guangzhou, China. Glow wire and needle flame apparatus are among our oldest product lines, with several hundred units in service across appliance, lighting and materials laboratories.
Our engineering team works directly from the IEC text and from customer samples. If your part is small, thick, ribbed or has a metal insert that changes heat flow, send the drawing and we will propose the clamping arrangement before you order.