| Brand Name: | Hongce |
| Model Number: | HC-225L-70 |
| MOQ: | 1 Set |
| Price: | Negotiation |
| Payment Terms: | T/T,L/C |
| Supply Ability: | 20 Sets per Month |
The TH series is a programmable climatic chamber used for temperature and humidity reliability testing of components, sub-assemblies and finished products. Chamber air temperature is controlled from -70 °C to +180 °C and relative humidity from 20 % to 98 % RH, so one unit covers cold storage, dry heat, steady-state damp heat and cyclic damp heat programmes.
The working volume is built in mirror-finish SUS304 stainless steel and insulated with 100 mm rigid polyurethane foam plus glass wool. A cascade refrigeration circuit with imported compressors provides the low-temperature range, while a PID controller with a 7-inch colour touch panel runs multi-step profiles of up to 120 segments and 999 loop repetitions, with the measured curve plotted in real time on screen.
Typical users are electronics, battery, automotive, appliance and material laboratories that need documented evidence of how a product behaves when the climate around it changes. The chamber supports the test procedures listed below, and the factory test report records actual measured values rather than catalogue figures only.
How the Chamber Is SpecifiedEvery quotation is built from three decisions: working volume, temperature and humidity range, and the interface options needed by the laboratory. The table below explains what each choice changes in practice, so the specification you receive matches the test programme you intend to run.
| Decision | Options | Practical Effect |
|---|---|---|
| Working volume | 80 L / 150 L / 225 L / 408 L / 800 L / 1000 L | Volume should be at least 3 to 5 times the volume of the test specimen group, so air can circulate around each item. |
| Temperature range | -70 °C to +180 °C (standard TH series) | Low-temperature capability depends on the cascade circuit; deeper cold or faster ramp rates are quoted separately. |
| Humidity range | 20 % to 98 % RH | Below 20 % RH a dry-air purge accessory is normally required; at the upper end, condensation control becomes the limiting factor. |
| Control and interface | 7-inch touch panel PID; RS485 / Ethernet / USB data export optional | Determines whether profiles are edited on the panel or from a PC, and how records are archived for audit. |
| Cooling and water supply | Air-cooled (standard) or water-cooled condenser | Water-cooled units are quieter and better suited to rooms with limited ventilation. |
| Parameter | Value |
|---|---|
| Model | TH-80L / TH-150L / TH-225L / TH-408L / TH-800L / TH-1000L |
| Temperature range | -70 °C to +180 °C |
| Temperature accuracy | ±0.5 °C (control) |
| Temperature uniformity | ±2.0 °C (no-load, measured per 9-point survey) |
| Heating rate | 3 to 5 °C/min (non-linear, no-load) |
| Cooling rate | 0.7 to 1.5 °C/min (non-linear, no-load) |
| Humidity range | 20 % to 98 % RH |
| Humidity accuracy | ±2.5 % RH |
| Controller | 7-inch colour touch screen, programmable PID, 120 segments, 999 cycles |
| Interior material | SUS304 stainless steel, mirror finish |
| Exterior material | Powder-coated steel plate; SUS304 exterior optional |
| Insulation | Rigid polyurethane foam and glass wool, 100 mm thickness |
| Refrigeration system | Cascade system; imported compressors (European origin) standard, alternative brands optional |
| Refrigerant | R404A and R23 |
| Observation window | Multi-layer tempered glass with built-in LED lighting |
| Power supply | AC 380 V ±10 %, 50 Hz, three-phase five-wire, or AC 220 V ±10 %, 50 Hz |
| Safety devices | Over-temperature, over-pressure, over-current, earth-leakage breaker, phase-sequence protection, water-shortage alarm |
The chamber is intended for use with the following procedures, and the manufacturer's test report references the clauses applied at commissioning:
| Standard | Procedure | Typical Setting |
|---|---|---|
| IEC 60068-2-1 | Cold | Specified low temperature, defined duration, specimen unpowered or powered |
| IEC 60068-2-2 | Dry heat | Specified high temperature with natural or forced air circulation |
| IEC 60068-2-30 | Damp heat, cyclic | 24 h cycles between 25 °C / 97 % RH and 55 °C / 93 % RH |
| IEC 60068-2-78 | Damp heat, steady state | Constant 40 °C / 93 % RH for the agreed duration |
| GB/T 2423 series | Equivalent national procedures | Same profiles as the corresponding IEC parts |
| MIL-STD-810 (selected methods) | Climatic methods | Applied where the customer's test plan refers to this standard |
Chamber performance is verified with a multi-channel data logger at nine positions inside the empty working volume, following the steady-state method used in most accredited laboratories. Where a customer needs third-party calibration, the unit can be prepared for on-site verification by an external body before shipment or after installation.
ApplicationsBefore packing, each chamber runs a continuous programme at the upper and lower temperature limits, and the recorded curve is kept with the unit's file. The shipment documents include the operation manual, electrical wiring diagram, spare-parts list, factory test report with the measured uniformity survey, and the calibration certificate issued at the factory. This documentation is normally what a laboratory needs when it prepares its own equipment qualification file.
Installation, Training and SupportInstallation is straightforward for a laboratory with a suitable power supply and floor space; remote video commissioning is included, and on-site installation with operator training can be quoted for larger orders or for sites where the electrical supply must be verified first. Spare parts such as door gaskets, humidity sensor elements and compressor protection modules are kept in stock for units already delivered.
Chambers carry a 12-month warranty on parts and labour, excluding consumable items and damage caused by incorrect power supply or unauthorised modification. Technical questions are answered by the engineering team that configures the units, so replies address the actual measurement setup rather than general product information.
Commercial Terms| Item | Details |
|---|---|
| Minimum order quantity | 1 set |
| Supply ability | 50 sets per month |
| Price range | USD 3,500 to 15,000 per set, depending on chamber size and configuration |
| Payment terms | T/T, L/C, Western Union |
| Delivery time | 15 to 25 working days after order confirmation; custom configurations quoted separately |
| Packaging | Wooden case with vacuum-formed foam supports, suitable for sea freight and air freight |
| Brand / origin | Hongce, Guangdong, China |
| Certification | CE, ISO 9001 |
Compare the total volume of the specimens you normally test. A useful guide is that the specimen group should fill no more than one third of the working volume, which leaves space for air circulation and keeps uniformity within the stated tolerance.
Can the chamber run a profile with both temperature and humidity steps?Yes. The PID controller stores combined temperature and humidity segments in one programme, which is the usual way damp heat cyclic procedures such as IEC 60068-2-30 are configured.
Is the cooling rate the same with a loaded chamber?No. Cooling and heating rates slow down as specimen mass and heat capacity increase. If your test plan specifies a rate under load, send the specimen details with the enquiry so the refrigeration capacity can be matched to it.
Which documents come with the unit?The operation manual, wiring diagram, spare-parts list, factory test report with the measured uniformity survey, and the factory calibration certificate. Third-party calibration can be arranged on request.
What information is needed for a quotation?Working volume, temperature and humidity range, the standard or internal test procedure you follow, power supply available at site, and any interface requirement such as RS485 or data export to a laboratory system. With these details a specification and price can be issued the same working day.
Tip:Product images are for promotional purposes only; please refer to the actual promotional materials for accurate details.