Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction protective conductor

Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction protective conductor

  • Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction protective conductor
  • Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction protective conductor
  • Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction protective conductor
  • Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction protective conductor
Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction protective conductor
Product Details:
Place of Origin: China
Brand Name: HongCe
Certification: Calibration certificate(cost additional)
Model Number: CT-1
Payment & Shipping Terms:
Minimum Order Quantity: 1 set
Price: Negotiable
Packaging Details: Plywood
Delivery Time: 20 days
Payment Terms: L/C, T/T
Supply Ability: 30 set per month
Contact Now
Detailed Product Description
Electrode Output Testing Voltage: AC40-50V Temperature Environment: 0-40
Test Sample: Showing Contact With The Relevant Parts Conforms To Standard: IEC60990
High Light:

iec tester

,

laboratory test equipment

Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction Protective Conductor

 

Standard:

 

IEC60990 figure 4 touch current and the measurement method of protective conductor current

 

Application:

 

This network is convenient to use and easy to operate, and the power supply switch is made separately, which can solve the shortcomings of power supply capacity of the previous machine and can test the more powerful products.

Easy Operation IEC60990 Figure 4 Touch Current Test Network For Perception Or Reaction protective conductor 0

RS 1500Ω
R1 10000Ω
RB 500Ω
C1 0.022μF
CS 0.22μF

 

Test sample:

 

mesuring network touch current weighted (sensory current or reaction current)

 

Preparation Before Test

 

The following apparatus are prepared well before testing.

 

a set of oscilloscope.

 

a set of 1times or 10 times voltage probe.

A set of power supply(the voltage or frequency can be adjusted or variable to the required test.).

 

Test Methods

 

1. Adjust the power supply to the voltage and frequency the test required, and then plug the input terminal of the contact current test switch into the power supply.(please do not electrify when the network is not connected.)

 

2. Insert the sample into the output terminal of the contact current test switch.

 

3. Connect the port B of contact current test switch with the port B of contact current testing network.

 

4. Connect the port A of the contact current test network to the place where the contact current is required. Such as connect the port A to the adapter’s output or shell when the network adapter is tested.

5. The Oscilloscope probe is clamped into U1 or U2 terminals according to the test requirements.

 

6. Connect the power supply, put switch “N” to the on position, put the positive phase to the positive phase position, switch line “G” to on position, check whether the tested sample is on work by normal electricity and is there a waveform on the oscilloscope.

 

7. According to the standard requirements after the sample is working, in the different mode of “positive phase”,“reverse”,“N line on”,“N line off”,“G line off”,“line G on”and so on, read voltage U1 and U2 directly by using oscilloscope,the contact current is equal to the peak value of U2 and then divide 500.

 

Announcements

 

During the test, it is necessary to pay attention to safety because you can touch some terminals.

 

According to GB/T 12113 and IEC 60990 standard requirements, the power supply must be in isolation, and zero line and ground wire must be connected together.(the simple way is to whether there exists voltage on zero line and ground wire, if voltage surpass 2V, such as 220V power, the different voltage between zero line and ground wire more than 100V, the zero line and ground wire must be connected.)

 

When testing, please pay attention to the polarity of the power supply. The fire line must be connected to the L end of the contact current test switch. If the connection is wrong, there is a danger of electric shock.

 

Connect the oscilloscope, test the sample, then connect the power after b-terminal line, and do not connect the test line when the power is connected.

 